Manufactured by Roche Diabetes Care – Important instructions on how to fit 2 separate key frames to prevent accidentally activating the pump (MDA/2019/009).
-
Aguettant Limited has informed the MHRA that there is an error in the Patient Information Leaflet (PIL) for batches of Atropine Sulfate 3mg/10ml Solution for injection in pre-filled syringe.
-
Manufactured by Medtronic Inc – a subset of dual chamber pacemakers may experience a loss of pacing therapy when programmed to a dual chamber mode with atrial-sensing (MDA/2019/008)
-
Actavis Group PTC EHF is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
Manufactured by Eurotrol B.V – may give readings below the values assigned to the product, leading to incorrect measurement results.
-
Macleods Pharma UK Limited is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
Manufactured by GE Healthcare – Device may stop ventilation when in PSV Pro Mode and users may be unable to change gas and agent settings when using End-Tidal Control.
-
Actavis Group PTC EHF is recalling the below batches from pharmacies as a precautionary measure due to possible contamination with N‑nitrosodiethylamine (NDEA).
-
List of field safety notices (FSNs) from medical device manufacturers from 17 - 21 December 2018
-
Manufactured by Acutronic Medical Systems AG – Ventilator may switch off without user input, deliver incorrect ventilation or may fail to alarm.
-
List of field safety notices (FSNs) from medical device manufacturers from 10 - 14 December 2018
-
Some valsartan containing products including certain batches supplied by Teva UK Limited and all unexpired batches supplied by Mylan are being recalled.
-
Philips M3538A lithium-ion batteries manufactured from 28 December 2017 to 20 March 2018 inclusive for the HeartStart MRx monitor/defibrillator may have an internal component failure.
-
Cadiasun Pharma GmBH has informed us that there is an error on the patient information leaflet for Caspofungin 70mg powder for concentrate for solution for infusion.
-
IPS Specials are recalling specific batches of Melatonin 10 mg capsules because they have been assigned an incorrect extended expiry date.
-
Kyowa Kirin is advising caution in use for one lot of Bleo-Kyowa®, Powder for Solution for Injection, 15,000 IU following the detection of glass particles (EL (18) A/09)
-
FDC Pharma are recalling batch 087B087 due to an out of specification result for an unknown impurity during stability testing. (EL(18)A/08)
-
Caution in Use: Distribute to Pharmacy and Clinic Level.
-
Summary List of field safety notices (FSNs) from medical device manufacturers from 26 February – 02 March 2018
-
Manufactured by AFP Medical – do not use affected nebulisers as they have been manufactured to unknown standards and their safety cannot be verified (MDA/2018/008).
-
Caution in Use - Distribute to Hospital Pharmacy, Ward, Chemotherapy Unit and Clinic Level. (EL (18) A/05)
-
Glaxo Wellcome UK Limited is recalling one lot of Seretide Accuhaler 50/250mcg because a manufacturing issue has resulted in a small number of units not delivering the full number of doses. (EL (18) A/04)
-
Glaxo Wellcome UK is recalling two Lots of Ventolin Accuhaler 200mcg because a manufacturing issue with these Lots has been identified which has resulted in a small number of units not delivering the full number of doses (EL…
-
Manufactured by ICU Medical (formerly Hospira) – potential for connectivity engine module to disengage from main chassis, which may lead to power down without an alarm (MDA/2018/006).
-
The supply of batteries for these devices has been temporarily stopped so there may be a significant delay in the fulfillment of orders for new batteries.
-
Caution in Use: Distribute to Hospital Pharmacy and Ward Level.
-
Caution in Use: Distribute to Pharmacy, Hospital, GP, Dental Practice, Ward and Clinic Level.
-
Summary List of field safety notices (FSNs) from medical device manufacturers from 27 November to 1 December 2017.
-
Summary List of field safety notices (FSNs) from medical device manufacturers from 23 to 27 October 2017.
-
Caution in Use: Distribute to Hospital Pharmacy, Ward, Chemotherapy Unit and Clinic Level
-
Caution in Use: Distribute to Hospital Pharmacy, Ward and Operating Theatre Level
-
Strathclyde Pharmaceuticals Ltd is recalling a single batch of Xarelto 20mg film-coated tablets (Rivaroxaban) because of reports of rogue blister strips of 15 mg tablets within two packs of 20 mg tablets.
-
(Focus Pharmaceuticals) Error with the barcodes on selected batches of 8 medicinal products – Class 4 Caution in Use (EL (17)A/08)
-
Torrent Pharma (UK) Ltd is recalling the below batches as a precaution because out of specification results for dissolution were obtained during routine stability testing.
-
Baxter Healthcare Limited are initiating a precautionary, company-led, ward-level recall of two batches of Sodium Chloride Infusion 0.9% w/v 1000 mL bags due to a potential leak with the bags.
-
Pharmacy and wholesaler level recall of two batches of Parallel Imported One-Alpha Drops x 10 ml because the dropper included in this pack delivers 0.05 micrograms of alfacalcidol instead of 0.1 micrograms as labelled on the…
-
Manufactured by Philips Healthcare – two battery related problems: 1) may fail to power on or reboots repetitively, 2) after unplugging from AC mains, may fail to deliver a shock or undertake pacing.
-
Manufactured by Physio-Control – risk of unexpected device shutdown due to an intermittent connection between the battery and device contacts.
-
All Alaris™ GS, GH, CC, TIVA, PK, Enteral Syringe Pumps, and, Asena™ GS, GH, CC, TIVA, PK, Syringe Pumps – risk of uncontrolled bolus of medicine
Manufactured by CareFusion/BD Medical – identify and replace broken backplate spring in the plunger assembly and note updated preventative maintenance schedule for these pumps
-
Manufactured by Dräger – ventilator may stop working if oxide layers accumulate due to the infrequent use of the control knobs (potentiometers).
-
Action within 48 hours: Novo Nordisk Ltd. is recalling seven batches of GlucaGen HypoKit due to a small number of needles being detached from the syringe.
-
CME has recommended the need for protection of these pumps when exposed to direct sunlight.
In addition, due to connector leakages, the manufacturer is recalling extension sets (specific product codes and batches) used with…
-
(CooperSurgical under different brand names) Risk of serious burn if device is used past its expiry date (MDA/2015/025)
-
(All manufacturers) Low risk of Mycobacterium infection in patients undergoing cardiac surgery, associated with heater-coolers used with cardiopulmonary bypass machines. Possible cause may be Mycobacterium-contaminated water…
-
(Boston Scientific) Rapid battery depletion leads to risk of loss of therapy. (MDA/2014/039)
-
(All manufacturers) Delays in acting on field safety notices can compromise patient safety (FSNs). (MDA/2014/037)
-
(Various manufacturers) Risk if alarms are ignored or damaged breathing systems are used. (MDA/2014/020)
-
(Medtronic) Risk due to incorrect selection and delivery of bolus amount. (MDA/2014/019)
-
(Pace Medical Inc) CE-marking suspended due to quality system failings. (MDA/2014/015)
-
This alert has been issued as B. Braun Melsungen AG has informed us that during regular visual inspection of retained samples of Sodium Bicarbonate solutions with bromobutyl stoppers precipitation in solution was observed. …