Attaching an incompatible needle-free connector (NFC) to an Ethypharm Aurum pre-filled syringe can block the syringe and prevent delivery of potentially lifesaving medicine. Only compatible NFCs should be used with pre-filled syringes from the Aurum range.
A summary of recent letters and notifications sent to healthcare professionals about medicines and medical devices, and information about the publication of two recent Device Safety Information pages.
B. Braun Medical Limited is recalling various product batches as a precautionary measure after traces of midazolam were detected in the batches listed in this notification.
List of Field Safety Notices (FSNs) from 21 to 25 August 2023.
Fresenius Kabi Limited have identified an error on the infusion bag packaged in certain batches.
BD have notified the MHRA that the sterility of some of their devices cannot be guaranteed due to quality issues with their third-party sterilisation provider.
Qualasept Ltd t/a Bath ASU is recalling a specific batch of Irinotecan 200mg/260mL in sodium chloride 0.9% w/v intravenous infusion and specific batches of Infliximab (Remsima) Intravenous infusion in Sodium Chloride 0.9%w/v due to risks of reduced product quality.
(Hospira) If pump is running on mains power and software detects that battery cannot be fully recharged, an alarm will sound and infusion will stop. (MDA/2013/006)
(Hospira) Due to pump faults. (MDA/2013/016)
The solvent used in the preparation of Rifadin infusion has been reformulated. As a result, it can only be diluted with either dextrose 5% solution or sodium chloride 0·9% solution
Novartis Pharmaceuticals has informed the MHRA that the solvent (water for injections in ampoules) co-packed with the impacted batches of Simulect powder for injection, may contain glass fragments approximately 20 – 800 µm in size.