Radon technical and laboratory services for professionals
Calibration, testing, validation and specialist measurement services for radon detectors, radon monitors and other equipment used by professional organisations.
The Radon Dosimetry Team operates a radon exposure facility for calibration, testing, and research purposes. Radon gas concentrations are measured using instruments that are calibrated to traceable standards using a source from an accredited laboratory.
Our expertise covers both passive and active (electronic) radon detectors, and our radon scientists have many years’ experience in applying that knowledge to more unusual measurement situations. We provide calibration of passive detectors, active radon monitors and radon integrated electronic devices.
Calibration services
Radon detectors must be calibrated in order to respond correctly. Passive detectors are calibrated before being issued. Electronic radon detectors must be calibrated before first use and annually thereafter if they are being used for compliance with the Ionising Radiations Regulations 2017.
Whether it is passive detectors, active instruments or one of the many Electronic Radon Integrated Monitors (ERIMs) introduced in recent years, these must all be calibrated to ensure that the radon monitor is giving an accurate and reliable reading.
Manufacturers of passive radon detectors can submit sample etched track detectors for known exposures to ensure that the detector elements are of the correct sensitivity and that their track reading system is functioning correctly.
For instruments being used to measure radon levels in workplaces, the Ionising Radiations Regulations 2017 apply. These regulations state that instruments must be calibrated before use and then annually.
We strongly recommend annual calibrations for any instrument being used to monitor radon.
This is a selective list of some of the calibration services we currently provide – if you don’t see your instrument, detector or device, contact Radon.Calibration@ukhsa.gov.uk.
Passive Detectors
Exposures in the Radon Chamber:
- batch of passive detectors exposed together, F not monitored [note 1]:
- up to 2000 etched track
- or up to 50 charcoal
- or up to 100 electret
- one week calibration of up to 2000 etched track monitors, F monitored
- longer exposures are available on a daily rate, F monitored
Note 1: we can accept more than one batch to be exposed at the same time.
Exposures in FRED:
- Up to 80 detectors [note 1]
Note 1: we can accept more than one batch to be exposed at the same time.
Active Instruments:
- Pylon or RDA plus 1 flow through cell or up to 5 passive cells or scintillation trays (additional trays or passive cells incur an additional charge)
- Alphaguard monitor
- Integrating Working Level Monitor (IWLM)
- Rad 7 or Rad 8 monitor
-
ERIMs:
- AER+ (Algade)
- Alpha E (Bertin instruments)
- Canary (now rebranded as the Corentium Home) (Airthings)
- Corentium Pro (Airthings)
- Sun Nuclear (Sunradon LLC)
- Radon Eye (Radon FTLab)
- Radon Scout (SARAD)
- Ramon (GTAnalytic SARL)
- Wave (Airthings)
For more information on the radon chamber or the Fast Radon Exposure Device (FRED), and the hiring of either, see Hire the Radon Chamber.
Do not send instruments or detectors for calibration until after you have contacted us and we have agreed a start date for the work.
For all enquiries, and to book a calibration, email Radon.Calibration@ukhsa.gov.uk.
We will provide you with a quotation, estimated turnaround time and potential date for commencement of the work.
Validate passive radon monitors
The validation scheme outlines requirements by which suppliers of passive radon detectors can demonstrate competence in the measurement of radon levels in UK dwellings and workplaces. The scheme includes a review of the quality procedures used by each supplier, performance tests that assess their ability to measure radon by exposing detectors to known quantities of radon, and guidance in the interpretation of the measured results.
Read about the validation scheme for measurements.
For enquiries or to apply your laboratory into the scheme, contact Radon.Calibration@ukhsa.gov.uk.
International intercomparison of passive monitors
Each year at the UKHSA Radiation, Chemical, Climate and Environmental Hazards Directorate (RCCE) laboratories in Oxfordshire, we carry out an international intercomparison of passive detectors for radon laboratories from around the world. The instruments in our 43 m³ Radon Chamber are calibrated to traceable standards using a source from an accredited laboratory. This means that we can provide reference exposures.
Sets of detectors from multiple laboratories are simultaneously given a range of undisclosed radon exposures and the objective is to determine those exposures with high accuracy and precision.
The passive detectors used for this purpose can be in the form of etched track detectors (PADC/CR-39, LR115, Makrofol®, etc), electret radon detectors, or (from UK laboratories only) activated charcoal detectors.
The detectors are randomised into 6 groups: 5 groups are exposed to different exposure levels in the range 100 to 2,400 kBq m⁻³, detectors from the 6th group are not exposed and record the exposures in transit.
There is a charge for each set of detectors submitted. A detailed report of the detector performance is supplied to the customer.
Following the annual intercomparison, an anonymised report of the findings from the participating laboratories is published. View the latest annual intercomparison of passive radon detectors report.
If you wish to take part in the next intercomparison, or require further information, contact Radon.Calibration@ukhsa.gov.uk.
Hire the Radon Chamber
The 43 m³, walk-in Radon Chamber is available for hire by customers to test equipment themselves, if necessary. Radon Group scientists will be available to adjust conditions within the chamber to the customer’s requirements (normally from 400 Bq m⁻³ to 8,000 Bq m⁻³). A small facility is also available with a radon concentration of around 80,000 Bq m⁻³.
The Radon Group at UKHSA and predecessor organisations has operated radon exposure facilities for over forty years, for research and the calibration of passive detectors and radon monitoring instruments. These facilities allow for exposures to known levels of radon gas that are measured using instruments which have been calibrated to traceable standards using a source from an accredited laboratory.
UKHSA Radon Chamber
The main exposure facility is a 43 m³ walk-in radon chamber that is accessed by an airlock. Radon gas is maintained at a sufficiently high level in the chamber to allow exposures to be completed in a relatively short time. For example, a set of passive detectors can be exposed to the equivalent of three months at the current UK Action Level of 200 Bq m⁻³ in a matter of days, while still allowing staff to safely enter the chamber.
Chamber details:
- radon concentration: 400 to 8,000 Bq m⁻³
- volume: 43 m³
- monitored continuously: radon gas, temperature, humidity, barometric pressure
- additional variables available for monitoring: decay products, working level, equilibrium factor, ambient aerosol
- controllable parameters: radon gas concentration, equilibrium factor (F), aerosol concentration
- uncertainties (95% Confidence Limit): radon gas 10%, decay products, working level, equilibrium factor 20%
The chamber is suitable for a wide variety of radon detectors and instruments. It has also been used to test other types of equipment, such as determining the radon response of actinide in air contamination monitors and etched-track neutron dosemeters.
Clients send their instruments or passive detectors to us for our staff to undertake the exposures. It is also possible for clients to carry out their own radon experiments by hiring the chamber on a daily basis. Monitoring, controlling and reporting the conditions in the radon chamber listed above throughout this time can be provided.
Fast Radon Exposure Device (FRED)
This facility has a much higher radon concentration compared to the radon chamber. This allows for either exposures in an even shorter amount of time, or much higher exposures, which can be useful when determining a detector’s linearity of response.
However, this facility is much smaller than the radon chamber, and so can only be used for a small number of detectors. It is too small for some active instruments, and any instruments placed inside would need their own inbuilt power source.
FRED details:
- radon concentration: 40 to 80 kBq m³
- volume: 0.2 m³
- monitored continuously: radon gas
- uncertainties (95% CL): radon gas 20%
If you are interested in hiring either of our facilities listed, contact Radon.Calibration@ukhsa.gov.uk.
Electronic radon integrating monitors (ERIMs)
ERIMs are electronic radon monitors that aim to determine accurately the long-term average radon concentrations. Following studies of ERIMs, we recommend certain helpful techniques. In addition, the detectors must be calibrated if they are to be used for regulatory compliance.
These detectors have gained popularity due to their relatively low cost and ease of use.
However, the visible display of a radon concentration on many of the ERIMs leaves them open to misuse. There is no such thing as an ‘instant’ radon reading as the physics of radioactive decay requires several days for any instrument to respond.
UKHSA studied the performance of some of these monitors recently and recommend the following:
-
instruments should be cleared of previous records and reset according to instructions – this often requires an extra step in addition to switching the instrument off and on again
-
before starting a measurement, the batteries should be replaced with new ones and/or the instruments should be fully charged
-
ERIM instruments require annual calibration if used for workplace measurements (when regulatory compliance is needed, e.g. the Ionising Radiations Regulations 2017)
-
instruments should be left undisturbed for at least few days before the data is downloaded or used (their low sensitivity means that they cannot respond quickly to a new location)
-
if ERIM instruments are intended for a long-term study of indoor radon, then they should follow the same instructions of placement as passive detectors
For more detail on the evidence supporting these recommendations, see Long-term comparison and performance study of consumer grade electronic radon integrating monitors.
Book a calibration for an ERIM, email Radon.Calibration@ukhsa.gov.uk.
We will provide you with a quotation, estimated turnaround time and potential date for commencement of the work.
Radon passive monitor services – harsh environments
Passive monitors can be used in unusual situations, provided that special provisions are made to ensure that they are protected from very high humidity, radioactive contamination and other harsh environments. These services are also suitable for measurements in mines, caves, tunnels, water treatment works, outdoors, and in gamma radiation and neutron fields.
As these environments are different from most other workplaces, there are multiple changes that are made to the monitoring protocol in terms of:
- placement
- duration
- seasonal corrections
- humidity
- contamination
- neutron radiation fields
- other considerations
Placement
The detectors need to be placed in locations that reflect the main working areas.
Mines: identify points in the main working areas and travel roads, taking the ventilation circuit into account along with cross-cuts and multiple levels. Old workings should be monitored if they are inspected for roof, etc, but no special access should be made for radon monitoring alone. There is no ‘correct’ number of detectors according to length of roadway or area/volume, although linking radon monitoring positions with those used for air flow measurements is helpful. The detectors should hang from cables or other fixtures at around head height; do not push them into crevices as this can adversely affect the results.
Caves: should be monitored in a similar way to mines. You may need to take pegs or similar items to fix the monitoring positions. If there are multiple entrances and exits to the cave system, detectors placed in these areas will indicate the average air flow.
Tunnels: ensure that a detector is placed securely at each end and at multiple points in between. If the tunnel is simple, few intermediate monitoring positions are required. Include side passages and refuges.
Water treatment works: can be monitored as for most ordinary buildings, ensuring that all main areas are represented.
Areas with radioactive contamination: can be monitored as for an ordinary building.
Areas with neutron radiation have two detectors placed at each location: one detector is standard; the second detector remains in a radon-proof bag.
Duration
Radon detectors placed in harsh environments can have their exposure time reduced to one month. This is to reduce the chances of the detectors being damaged and because the seasonal variations in radon levels will not be typical. Mines and caves can have extreme variations between summer and winter, with the summer showing the highest radon levels (in contrast to most ordinary buildings).
Seasonal corrections
As these are not typical buildings, we cannot estimate the annual average radon concentration from the typical seasonal variations. We recommend that you anticipate two monitoring periods in opposite seasons to obtain a pair of results for each location, which can be averaged to obtain a value that can be compared to the Ionising Radiations Regulations 2017 threshold of 300 Bq m⁻³.
Humidity
High relative humidity can prevent a result being obtainable as condensation forming inside the radon detectors forms a protective film over the detecting element. To prevent this, radon detectors can be issued in water-resistant/radon-permeable bags (1, 2, 3). These must be requested at the time of ordering and the detectors must not be removed from the bags, otherwise the results will be invalid.
Contamination (radioactive)
Radon detectors can be placed in areas with loose radioactive contamination, using the water-resistant/radon-permeable bags. The detector must stay in the bag during exposure. The outside of the bag must be checked by Health Physics/RPS/RSO before being returned for processing. If contamination is found that cannot be removed, then the detector can be taken from the bag and checked further, then returned for processing if clear. This must be clearly marked on the record sheet. A clearance certificate from Health Physics should be returned with the detectors.
Neutron radiation
Radon detectors are sensitive to neutron radiation (2). In order to differentiate the signal (damage tracks on the CR-39 element) from the radon and the neutrons, 2 monitors are exposed side-by-side. One detector records both the radon and neutrons; the other – in a radon-proof bag – records only the neutrons. Subtracting the latter from the former produces only the radon exposure for that location. The detector in the radon-proof bag must remain in the bag during and after exposure.
Gamma radiation
Gamma radiation cannot be detected by radon detectors. However, high gamma fields can cause structural damage to the CR-39 element, which makes it unable to record radon exposure properly. The background gamma radiation is usually too low to cause such damage (2). If radon levels are to be measured in the presence of high gamma fields, contact Radon.Calibration@ukhsa.gov.uk for advice.
UV and sunlight
The radon detector’s sensing element is very sensitive to sunlight and UV radiation. It is normally well protected from exposure by the outer casing, however, placing the detector directly in sunlight should be avoided to eliminate infrared (thermal) radiation that can greatly influence its ability to detect radon correctly (2,3,4).
Email Radon.Calibration@ukhsa.gov.uk if you would like to test in any of the above environments.
Measurement of radon in natural gas samples
The Radon Group has facilities for testing samples of natural gas for radon gas content. Radon levels are assessed in relation to the requirements of the Radioactive Substances (Natural Gas) Exemption Order 2002, statutory Instrument No. 1177.
The samples must be supplied to our laboratory in appropriate containers, to prevent combustion of the gas. The activity of radon in the sample is measured using a Pylon AB-5 monitor and scintillation cell with brass connectors. The sample container type, collection and delivery to our laboratory at Chilton must be arranged in advance.
Email Radon.Calibration@ukhsa.gov.uk if you would like to have any measurements of natural gas samples performed.
Testing of actinide in air monitors
There is often a general requirement that actinide in air monitors do not respond to radon and its decay products to prevent the occurrence of false positives or alarms. The Radon Chamber can be hired to allow the customer to demonstrate that their monitors do not respond to radon.
Email Radon.Calibration@ukhsa.gov.uk if you would like to like to learn more.
References
- Miles J, Ibrahimi F and Birch K (2009) ‘Moisture-resistant passive radon detectors’, Journal of Radiological Protection, volume 29, issue 2, page 269
- Wasikiewicz JM (2018) ‘Impact of environmental factors on PADC radon detector sensitivity during long term storage’, Radiation Physics and Chemistry, volume 142, pages 141 to 145
- Homer JB and Miles JCH (1986) ‘The effects of heat and humidity before, during and after exposure on the response of PADC (CR-39) to alpha particles’, International Journal of Radiation Applications and Instrumentation Part D: Nuclear Tracks and Radiation Measurements, volume 12, issues 1 to 6, pages 133 to 136
- Hussain G (1982) ‘Influence of ultraviolet radiation on the nuclear track recording properties of CR-39 polycarbonate detectors’, International Journal of Applied Radiation and Isotopes, volume 33, issue 7, pages 517 to 519