Biological safety assessment
Published 30 March 2026
This guidance sets out the information that the MHRA expects to see with a clinical investigation submission so that we can verify that the requirements of the relevant regulations have been met.
A biological safety evaluation must be carried out before any clinical investigation begins to ensure you comply with Part II of the UK Medical Devices Regulations 2002 (SI 2002 No 618, as amended) (UK MDR 2002), Annex VIII (as modified by Part II of Schedule 2A to the UK MDR 2002), and Annex XV of the (EU) Regulations for Medical Devices 2017/745 (EU MDR). This is to provide an acceptable level of reassurance to the patients in a clinical investigation, but also because biological safety is an aspect of conformity assessment that will never, in its entirety, be investigated by the proposed clinical study.
An assessment of toxicological risks is necessary for the assurance of biological safety. Other biological risks, such as microbial contamination, are excluded from this type of assessment.
The manufacturer of a device is responsible for documenting the results of the biological safety evaluation. In assessing a notification, the MHRA must be able to verify that an appropriate toxicological risk assessment has been carried out so that we can ensure that the anticipated benefits to the patients enrolled in the clinical study justify the foreseeable risks (Part II of the UK MDR 2002, Annex VIII, Section 3.2 [as modified by Part II of Schedule 2A to the UK MDR 2002]) and (Article 62 of the EU MDR).
The final responsibility for the safety of the product remains with the manufacturer but sufficient data should be submitted to the MHRA for review to provide assurance that all necessary toxicological risks have been appropriately considered and mitigated in line with ISO 14971:2019.
Conformity assessment requirements
Several of the essential requirements in the UK MDR 2002 and the general safety and performance requirements set out in Annex I of the EU MDR refer to, and apply to, the biological safety of the device. They describe which aspects should be taken into account in the design and manufacture of a medical device. The main safety aim is that the device will not compromise the clinical condition or safety of the patient or user or other persons ‘provided that any risks which may be associated with their use constitute acceptable risks when weighed against the benefits to the patient…’.
Qualified personnel are needed to perform a state-of-the-art risk benefit analysis to determine what are acceptable risks. If such expertise is not present within the company, external advice from suitably qualified sources should be sought. The requirements regarding design and manufacture of a medical device provide some detail as to what aspects of biological safety should be considered, including:
- the choice of materials used as regards toxicity (UK MDR 2002 Requirement 7.1 or EU MDR Requirement 10.1)
- the compatibility between the materials used and biological tissues, cells and body fluids, taking account of the intended purpose of the device, and where relevant, absorption, distribution, metabolism and excretion (also UK MDR 2002 Requirement 7.1 or EU MDR Requirement 10.1)
- minimisation of risk from contaminants and residuals, where particular attention must be paid to the tissues exposed and the duration and frequency of exposure (UK MDR 2002 Requirement 7.2 or EU MDR Requirement 10.2) - the concept minimisation rather than acceptable risk is used here because residuals and contaminants are expected to only bring risk and no benefits
- safe use of the device with the substances they will enter into contact with during their normal use (UK MDR 2002 Requirement 7.3 or EU MDR Requirement 10.3)
- risks from substances, including wear debris, that have the potential to be released from the device must be minimised - inclusion of substances within a medical device that are classified as carcinogenic, mutagenic, or toxic to reproduction (CMR) or substances whose endocrine disrupting properties provides risk to human health, above 0.1% weight by weight must be appropriately justified. (EU MDR Requirement 10.4.1)
- risks from unintentional ingress from substances into the device must be minimised (UK MDR 2002 Requirement 7.6 or EU MDR Requirement 10.5)
- minimise the risks posed by nanomaterials (EU MDR requirement 10.6)
- risks from ageing of the materials used in situations where the device cannot be maintained or calibrated (such as implants) should be minimised (UK MDR 2002 Requirement 9.2 or EU MDR requirement 14.2) - from a toxicity point of view, this means that breakdown products of the materials used should be taken into account in the risk assessment of long-term devices
- use of substances that have allergic potential, particularly the ability to induce type 1 allergy should be properly risk assessed and appropriate risk mitigation steps taken such as clearly identified with suitable warnings
Toxicological risk assessment
This assessment should include an evaluation of all relevant toxic endpoints according to the device’s nature and anticipated contact with human tissues when in use. A useful matrix of such endpoints exists in the standard ISO 10993-1:2018. Please note that all recommended endpoints are required to be sufficiently assessed and adequately documented.
Recommended endpoints for consideration should not be interpreted as a list of required biological tests to be carried out. This point is made clearer in the current version of the standard where “endpoints” to be considered are now denoted with an “E” for evaluate rather than previous versions having “X” often misinterpreted as standing for tests.
Physical and chemical information on the medical device is now included in the matrix to explicitly inform users of the standard that any biological risk assessment should start with full material characterisation of the final medical device. In all cases a complete chemical make-up of the final medical device after all raw materials have gone through the manufacturing process should be provided and such chemicals undergone sufficient toxicological risk assessment.
If insufficient chemical information exists for the final device to adequately carry out a risk assessment, data generation using appropriate analytical chemistry testing should be considered in accordance with ISO 10993-18:2020. The extent of physical characterisation required will be dependent on the medical device in question and is particularly important in the assessment of implants and blood contacting devices.
It should be kept in mind that individual considerations of the device might lead to a need to assess toxic end points over and above the ones listed in the matrix. Such ‘individual considerations’ include, but are not limited to the site of exposure of the device and the duration of exposure. For example, regarding exposure duration, long-term exposure can be obtained via a single device application (such as implants), or via repeated use of new single-use devices (such as contact lenses, colostomy bags). In the first case, the effects of degradation products need to be assessed in addition to the exposure to the device materials. In the second case, long-term exposure to the device materials still needs to be assessed, but individual device exposure might be short enough that the exposure to breakdown products could be negligible.
The guidance on the specific tests in different parts of the ISO 10993 series only discusses the test and sample methods, it does not specify pass or fail criteria. It is not possible to comply with the essential requirements or general safety and performance requirements by listing the tests done according to ISO 10993 and indicating a ‘pass’ behind each test. The results of the tests should be interpreted in the context of the overall risk assessment to know whether a specific outcome indicates acceptable risk or not. This once again emphasises the need for an overall, scientifically valid risk assessment carried out by suitably qualified individuals.
The three basic types of information required for a toxicological risk assessment are identified in Annex C of the standard BS EN ISO 14971:2019, for example:
- the physical and chemical nature of the materials (including the toxicity of ingredients and chemicals introduced during the manufacturing process)
- prior use of the materials
- biological safety test data
Toxicological hazard is a property of the chemical constituents of the materials from which a medical device is made and should be considered in relation to the hazard identification steps of BS EN ISO 14971. The second and third types of data listed above are dependent upon exposure to a particular material or device and thus predominantly address the risk of an adverse reaction occurring, rather than hazard.
Hazard identification
The toxicity of a compound or material is investigated using basic toxicology studies that provide information on the nature of toxic effects elicited by a test material, the dosages at which the effects occur and the no-effect levels. There is a widespread misconception that ‘medical grade’ materials (or ‘surgical grade’ or ‘implant grade’) exist. In the UK, there is no regulation or standard that defines what a medical grade material would be and indeed, considerations to achieve acceptability for implantation in an orthopaedic implant would be quite different for a vascular stent.
Compliance of a material with a pharmacopoeial requirement does confer some reassurance insofar as that is a well-recognised starting point. BS EN ISO 14971 requires that relevant characteristics that could affect safety are listed and, from these, possible hazards are identified. For a biological safety assessment, the first step comprises characterisation of materials. Data relating to formulations, additives, processing aids, degradation products, effects of processing, residue levels and similar factors allow the chemical nature of the materials to be fully characterised. Annex B of ISO 10993-1:2018 contains guidance in the event that the necessary data is not available to a manufacturer due to confidentiality or proprietary information.
Toxicological hazards can be identified from a knowledge of the toxicity of the chemicals or materials listed.
Risk evaluation
The probability that an adverse effect will arise from exposure to a chemical depends not only on its inherent toxicity, but also on the amount to which a participant is exposed and the route of exposure. This means that information on the identity and toxicity of materials, residues, degradation products and similar factors must be considered in relation to quantitative data on the amount present in the final product, its leachability and systemic availability. This analysis would fulfil the requirements of Clause 5.5 of BS EN ISO 14971:2019. In practice this means that, in many cases, an exposure assessment will already provide sufficient justification for limiting the number of toxic end points that need to be addressed in more detail. Where significant risks arising from hazardous residues are identified by this procedure, their acceptability should be assessed in line with established toxicological principles.
Additional information from biocompatibility tests or prior use of materials may be used to provide a basis for further assessment of risks. For prior use data, once again the site and duration of exposure of the new device compared to the prior use should be kept in mind. These data provide either an estimate of risk in line with Clause 5.5 of BS EN ISO 14971:2019 or a measure of the acceptability of risks in line with Clause 6, respectively.
Acceptable results from appropriate biological tests, of the sort described in the ISO 10993 series of standards, may give a degree of reassurance that the risk of adverse reactions occurring during clinical use is low. ISO 10993-1:2018 indicates evaluation tests which can form a framework for the development of an assessment programme. These tests, commonly termed biocompatibility tests, differ from basic toxicity tests in that they typically attempt to mimic the conditions of clinical exposure to medical devices and thus provide an indication of the probability of adverse effects arising during use. They tend, as a result, to be less sensitive than basic toxicity tests and are a less discriminating indicator of risk. Biocompatibility test data should therefore be used to complement an assessment based on materials characterisation, rather than as a replacement for it.
It may not be necessary to perform tests suggested by the standard in all cases. Where the proposed materials have already been extensively used or tested, little or no biological testing may be necessary to judge the biological safety of the device. Conversely, the need for additional tests such as mutagenicity or metabolism studies should be considered on a case-by-case basis. National laws may prohibit unnecessary animal tests and the need for testing should always be judged in relation to the predictive value of the test and animal welfare considerations.
Well established and documented clinical use of materials or their chemical components in an analogous situation can provide an indication of their suitability for an intended purpose. Risks anticipated from the intended use are assumed to be similar to those known, through clinical experience, to be acceptable in a comparable situation. It is common practice to evaluate biological safety primarily on the basis that the materials have been used before in similar applications.
This sort of evaluation requires sufficient knowledge of the chemical identity of the material or chemical to determine equivalence. It also requires clinical evidence to demonstrate that the material or chemical has performed with a suitable biological response. This may be in the form of the results of a clinical investigation or data relating to previous commercial use and adverse incidents. It will be clear that the absence of adverse effects in commercial use without some feedback mechanism to identify potential problems is of very limited value.
It should be noted that for chronic toxic endpoints, such as mutagenicity, this approach generally provides less reassurance as it is unlikely that enough accurate epidemiological data are available to link adverse effects with a long incubation period (such as cancer) with the use of certain products. As mentioned before, differences in sites and durations of exposure should be taken into account and might significantly reduce the usefulness of the prior use data. It is important to recognise that, even where an assessment is based entirely on previous clinical experience, the biological safety evaluation needs to be documented by the manufacturer and that such an evaluation is just as much in line with the principles set out in ISO 10993-1 as one that incorporates biological testing.
Documentation of a biological safety assessment
Sufficient pre-clinical data must be documented by the manufacturer to demonstrate how the biological safety of the device was assessed. The amount and nature of data necessary for an assessment varies with the intended nature and duration of patient contact and with the nature of the data that are readily available. It is not possible to identify precisely what information may be needed in any given instance since deficiencies in one area may be compensated for by data of another type. In general, biological safety data should be assessed in respect of all components with tissue, blood, mucous membrane or fluid path contact.
Documentation of an evaluation for the purposes of a manufacturer’s conformity assessment should normally include or reference the following:
- a general description or drawing for the device
- quantitative formulations for all materials in contact with tissues or infusible fluid
- a review of available toxicity and prior use data for each material and chemical in contact with tissues or infusible fluid
- information on previous experience with each material in related applications
- reports of biological safety tests
- an assessment of the data reviewed
- CV of the people responsible for, and contributing to the documentation of, the biological evaluation
What could be included in your documentation
A general description or drawing of the device
For each component, the generic (for example, PVC) and trade names of the materials used, the producer of the material, the nature of patient contact (for example, blood, fluid, topical or non-contact) and duration of proposed use should be stated.
Quantitative formulations for all materials in contact with tissues or infusible fluid
This should comprise the chemical names and concentrations of all ingredients, including additives, processing aids and catalysts. Where the presence in the final product of biologically active ingredients or reaction products is possible, the need for information on residue concentration should be considered. Where exposure to the device will be long-term, the potential for in situ degradation and the identity of the potential degradation products should be considered.
Where formulation data, or any other information relevant to the clinical investigation submission, are not available to a manufacturer, arrangements should be made for the supplier to make these available to the MHRA on a confidential basis. Where data are excluded from the documentation due to confidentiality, alternative methods such as chemical testing should be considered.
A review of available toxicity and prior use data for each material and chemical in contact with tissues or infusible fluid
This should include, where appropriate, data on residual monomers, contaminants, additives, catalysts, mould release agents, sterilisation agents and other processing chemicals and degradation products.
Data should be sufficient either to characterise the toxicity of the material or to indicate that it can be considered suitable for the proposed use. Where significant residues of biologically active ingredients are present, the need for data on their migration, leaching or absorption should be considered. Toxicity information may not be necessary for materials or chemicals which have been widely used in medical devices with similar exposure conditions and a documented history of absence of adverse effects. Justification for such omissions should be given.
Information on previous experience with each material in related applications
Where available, a summary of experience from clinical use of the materials should be documented. This should include details of the nature of patient contact, the number of devices used, the period of time for which they have been used and the number and nature of adverse reactions. For novel materials, this information may be necessary for each ingredient. Where the mode of patient contact or the chemical or physical nature of the established material is not identical to that in the intended application, the differences should be identified and an assessment of their significance documented.
Where relevant, the results of clinical investigations which address biological safety should be included.
Reports of biological safety tests
The full reports of biological safety tests carried out on the device as a whole or on component materials should be available. A scientific justification for the performance of each test, or the omission of any test suggested by 10993-1:2018, should be documented.
An assessment of the data reviewed
The documentation should include an appraisal of the toxicological significance of the data, by someone with experience in the assessment of the biological safety of medical devices. Suitability of the materials for the intended use should be judged on the basis that there is sufficient information to provide a realistic level of assurance that the risk/benefit ratio is acceptable or toxicological risks are no greater than those currently deemed acceptable for existing devices.
Further information
If you have any questions about biological safety assessments, or the clinical investigation procedure, you can email us at info@mhra.gov.uk.