Medical safety alert

Tianjie Dental Falcon portable X-ray unit and other non CE-marked dental X-ray units - lacks sufficient shielding in the X-ray tube

Identify and stop using this and similar non CE-marked devices. (MDA/2012/046)

CAS deadlines

Action underway: 23 July 2012, action complete: 06 August 2012

Device

Medical Device Alert Non CE-marked portable dental X ray units including the Tianjie Dental Falcon

Non CE-marked portable dental X-ray units.

Including the Tianjie Dental ‘Falcon’.

The Tianjie Dental Falcon portable X-ray unit is not CE-marked as a medical device and was purchased on eBay. It is identical in appearance to a number of other non CE-marked devices that are available on the internet:

Medical Device Alert Non CE-marked portable dental X ray units including the Tianjie Dental Falcon

Problem

Testing by the Health Protection Agency (HPA) has revealed that the Tianjie Dental Falcon device lacks sufficient shielding in the X-ray tube, which could give rise to high patient doses and under typical high radiographic workloads result in operator doses in excess of the Ionising Radiation Regulations 1999 annual dose limits. This could give rise to adverse health effects caused by radiation.

The Tianjie Dental Falcon is not CE-marked as a medical device.

Action

Action by

Dentists, Medical Physics Departments

Distribution

This MDA has been sent to:

  • NHS trusts in England (chief executives)
  • Care Quality Commission (CQC) (Headquarters) for information
  • HSC trusts in Northern Ireland (chief executives)
  • NHS boards in Scotland (equipment coordinators)
  • Local authorities in Scotland (equipment coordinators)
  • NHS boards and trusts in Wales (chief executives)
  • Health and Safety Executive
  • Primary care trusts in England (chief executives)

Onward distribution

Please bring this notice to the attention of relevant employees in your establishment.  Below is a suggested list of recipients.

Trusts
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:

  • Dental Hospitals
  • Dentists
  • Orthodontists
  • Medical physics departments
  • Radiographer superintendents
  • Radiographers
  • Radiologists
  • EBME Departments
  • Supplies departments

Primary care trusts
CAS liaison officers for onward distribution to all relevant staff including:

  • Community dental practices
  • Directors of public health
  • General dental practitioners
  • Maintenance staff

Independent distribution

Establishments registered with the Care Quality Commission (CQC) (England only)
This alert should be read by:

  • Hospitals in the independent sector
  • Dental Practices

Please note: CQC and OFSTED do not distribute these alerts. Independent healthcare providers and social care providers can sign up to receive MDAs directly from the Department of Health’s Central Alerting System (CAS) by sending an email to: safetyalerts@dh.gsi.gov.uk and requesting this facility.

Manufacturer contact

Lin Lin
Zhengzhou Tianjie Electronic Equipment Co
No 3 Youai Road
Zhengzhou City
Henan Province
P.R. China

Tel:    +86 371 67934274
Fax:   +86 371 67375396

website: http://www.tianjiedental.com/contact.html

Feedback

If you have any comments or feedback on this Medical Device Alert please email us at: dts@mhra.gsi.gov.uk

Appendix

Appendix: Non CE-marked portable dental X-ray units including the Tianjie Dental ‘Falcon’(MDA/2012/046) (622Kb)

England

If you are in England, please send enquiries about this notice to the MHRA, quoting reference number MDA/2012/046 or 2012/006/025/081/008.

Technical aspects

David Grainger or Paul Sandhu
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ

Tel: 020 3080 7199 or 7266
Fax: 020 8754 3965

Email: david.grainger@mhra.gsi.gov.uk or paul.sandhu@mhra.gsi.gov.uk

Clinical aspects

Jonathan Plumb
Medicines & Healthcare products Regulatory Agency
Floor 4
151 Buckingham Palace Road
London SW1W 9SZ

Tel: 020 3080 6648
Fax: 020 8754 3965

Email: jonathan.plumb@mhra.gsi.gov.uk

How to report adverse incidents

Please report via our website: Reporting adverse incidents involving medical devices

Further information about CAS can be found on the CAS website

Northern Ireland

Alerts in Northern Ireland will continue to be distributed via the NI SABS system.

Enquiries and adverse incident reports in Northern Ireland should be addressed to:

Northern Ireland Adverse Incident Centre
Health Estates Investment Group
Room 17
Annex 6
Castle Buildings
Stormont Estate
Dundonald BT4 3SQ

Tel: 02890 523 704
Fax: 02890 523 900

Email: NIAIC@dhsspsni.gov.uk

How to report adverse incidents in Northern Ireland

Please report directly to NIAIC, further information can be found on the NIAIC website

Further information about SABS can be found on the SABS website

Scotland

Enquiries and adverse incident reports in Scotland should be addressed to:

Incident Reporting and Investigation Centre
Health Facilities Scotland
NHS National Services Scotland
Gyle Square
1 South Gyle Crescent
Edinburgh EH12 9EB

Tel: 0131 275 7575
Fax: 0131 314 0722

Email: nss.iric@nhs.net

Health Facilities Scotland Incident Reporting and Investigation Centre (IRIC)Health Facilities Scotland Incident Reporting and Investigation Centre (IRIC)

Wales

Enquiries in Wales should be addressed to:

Improving Patient Safety Team
Medical Directorate
Welsh Assembly Government
Cathays Park
Cardiff CF10 3NQ

Tel: 029 2082 3922

Email: Haz-Aic@wales.gsi.gov.uk

Download documents

Medical Device Alert - Non CE-marked portable dental X-ray units including the Tianjie Dental ‘Falcon’(MDA/2012/046) (346Kb)

Help us improve GOV.UK

Don't include personal or financial information, eg your National Insurance number or credit card details.